Agilent 1290 Infinity LC UHPLC System Application Note for Triglyceride Analysis

This application note describes a method for high-resolution analysis of intact triglycerides under non-aqueous reversed phase gradient conditions using the Agilent 1290 Infinity LC System.


Agilent 1290 Infinity LC UHPLC System Application Note for Triglyceride Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
321 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
399507FDABD47A4F472530CE46DD874C

About this manual

The document details a method for analyzing intact triglycerides from vegetable and animal sources using reversed-phase HPLC under non-aqueous conditions. The study focuses on maximizing resolution for soybean oil, which is significant for biofuel production.

The analysis utilizes the Agilent 1290 Infinity LC System equipped with ultraviolet/visible (UV/VIS) Diode Array detection. Key experimental parameters include the use of C18 columns with high pressure configurations (up to 1200 bar), mobile phases consisting of acetonitrile mixed with isopropanol or methyl tert-butyl ether, and specific gradient elution profiles.

What's inside

  • Introduction
  • Sample Preparation
  • LC Method Details
  • Results and Discussion
  • Chromatographic Conditions

Frequently asked questions

What is the maximum operating pressure capability of the Agilent 1290 Infinity LC System?

The system is capable of operating at pressures up to 1200 bar, which is approximately 18,000 psi.

What column temperature was used for the analysis shown in Figure 4?

The column was prepared for operation at a temperature of 20 °C.

What is the maximum operating pressure recorded during the analysis in Figure 4?

The maximum observed pressure reached 760 bar.