Agilent 7700x ICP-MS ISIS Discrete Sampling Application Note

This Application Note describes how to use the Agilent 7700x ICP-MS with ISIS Discrete Sampling to perform high-throughput analysis of high-matrix samples. Using ISIS-DS, difficult, high-matrix analyses can be performed in a fraction of the time typically required, with no compromise in analytical performance.


Agilent 7700x ICP-MS ISIS Discrete Sampling Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
1.3 MB
Published
02 July, 2012
Updated
30 September, 2023
Model
7700x
MD5 checksum
1DECC1FBE6FDD54FAC3BEA1F13B2CCBD

About this manual

The document describes a method to maximize productivity when analyzing difficult, high-matrix environmental samples such as soils and sludges. By combining helium collision cell mode, aerosol dilution using the High Matrix Introduction (HMI) system, and ISIS Discrete Sampling (ISIS-DS), the Agilent 7700x ICP-MS can perform robust analyses in a fraction of the time typically required.

What's inside

  • Introduction
  • Instrumentation
  • Initial Calibration
  • Results
  • Method Detection Limits
  • Meeting EPA QA/OC Requirements

Frequently asked questions

How much time does it take to analyze a single sample?

The system performs a run-to-run analysis in 1.99 minutes per sample.

How many total analyses can be completed in an eight-hour sequence?

A total of 275 analyses were performed in 8 hours, 8 minutes.

What is the required recovery percentage for internal standards?

Internal standard recoveries must exceed 70% of the ISTD response in the calibration blank.

Download PDF (1.3 MB)

More Agilent Manuals


View all Agilent manuals →