This application note demonstrates a single quad GC/MS analysis of the GC/MS amenable listed allergens using the unique selectivity of an Agilent J&W DB-XLB capillary GC column
The document describes a method to detect 24 European Union (EU) regulated flavor and fragrance allergens in essential oils used for aromatherapy. It highlights the use of Gas Chromatography-Mass Spectrometry (GC/MS) coupled with an Agilent J&W DB-XLB capillary column, which offers unique selectivity to resolve these analytes from complex oil matrices.
The note details the experimental setup, including specific chromatographic conditions and flow path supplies used on an Agilent 7890A GC System. It presents results showing the detection of allergens in ylang ylang, lavender, and eucalyptus oils, demonstrating that the method can identify potential allergens at levels significantly higher than typical cosmetic limits.
| Column Dimensions | 30 m x 0.25 mm x 0.25 pm |
|---|---|
| Column Part Number | 122-1232 |
| GC System Model | 7890A GC System |
| MSD Model | 5975C series GC/MSD |
| Detector Type | Triple Axis Detector |
| Injection Volume | 1.0 pL |
| Split Ratio | 50:1 split ratio |
Twenty-four of the 26 EU-regulated flavor and fragrance allergens can be analyzed using GC/MS.
The analysis uses an Agilent J&W DB-XLB capillary GC column.
The oven starts at 50 °C, ramps to 100 °C, then 110 °C, and finally reaches 280 °C.