Agilent 7000 Series Triple Quadrupole GC/MS Application Note for Phytoestrogen Analysis


Agilent 7000 Series Triple Quadrupole GC/MS Application Note for Phytoestrogen Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.5 MB
Published
02 July, 2012
Updated
02 July, 2012
MD5 checksum
FFA5238BB6E4A509A25884D520EBC5DC

About this manual

This document details an analytical method developed for the identification and quantitation of eight plant phytoestrogens (biochanin A, coumestrol, daidzein, equol, formononetin, glycitein, genistein, and prunetin) in soy milk. The method utilizes gas chromatography tandem mass spectrometry (GC/MS/MS) with an Agilent 7000 Series Triple Quadrupole GC/MS instrument.

The procedure involves extracting phytoestrogens from soy milk using liquid-liquid extraction with ethyl acetate, followed by derivatization to trimethylsilyl ethers using trimethylchlorosilane (TMCS) and N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA). The analysis employs specific Multiple Reaction Monitoring (MRM) transitions optimized for each analyte to ensure high selectivity and sensitivity.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation and Derivatization Procedure
  • GC/MS/MS Instrumentation
  • Optimization of GC/MS/MS Conditions
  • MRM Transitions
  • Fragmentation Patterns

Specifications

Column TypeAgilent J&W HP-5 (5% phenyl, 95% methylpolysiloxane)
Column Dimensions30 m x 0.25 mm id
Carrier GasHelium
Flow Rate1.2 mL/min
Injector Temperature280 °C
Ion Source Temperature300 °C
Electron Energy-70 eV
Dwell Time50 msec

Frequently asked questions

Which phytoestrogens were analyzed in this study?

The study analyzed eight phytoestrogens: biochanin A, coumestrol, daidzein, equol, formononetin, glycitein, genistein, and prunetin.

What solvent was used for the extraction of phytoestrogens from soy milk?

Ethyl acetate was used as the extracting solvent for the liquid-liquid extraction of phytoestrogens from soy milk.

What reagents were used for the derivatization of the analytes?

The analytes were derivatized using trimethylchlorosilane (TMCS) and N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA).

What software was used for instrument control and data analysis?

MassHunter software was used for instrument control and data analysis.

Download PDF (1.5 MB)