Agilent 6410A Triple Quadrupole LC/MS Application Brief for Chloramphenicol Analysis

This document describes the use of LC/MS/MS 6410 for the analysis of chloramphenicol, thiamphenicol, and florfenicol in fish samples. The document also mentions that these compounds are banned in the EU and that LC/MS/MS is used for its greater sensitivity and higher selectivity. The method allows for the detection of all three compounds within less than 6 minutes and shows good performance in negative mode.


Agilent 6410A Triple Quadrupole LC/MS Application Brief for Chloramphenicol Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
86 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
6410A
MD5 checksum
6E63C78D41B44FE10D347AEA881D4BDA

About this manual

This document presents an application method for analyzing chloramphenicol, thiamphenicol, and florfenicol in fish samples using an Agilent 6410A triple quadrupole LC/MS system. The method utilizes a simple reversed-phase liquid chromatography (RRLC) approach to separate the compounds within less than 6 minutes, demonstrating high sensitivity in negative mode.

The procedure details sample preparation protocols for honey, shrimp, and chicken, including solid-phase extraction (SPE) steps. It also provides specific MRM settings for the analysis and reports linearity and sensitivity results, showing signal-to-noise ratios for the target compounds at low concentrations.

What's inside

  • Chloramphenicol Analysis
  • Thiamphenicol and Florfenicol
  • Fish Sample Preparation
  • LC Conditions
  • MSD Conditions
  • Sample Preparation (SPE)
  • MRM Settings
  • Sensitivity Results

Specifications

ColumnAgilent ZORBAX Eclipse Plus, 2.1 mm x 50 mm, 1.8 pm
Flow rate0.4 mL/min
Column compartment temperature45 °C
Injection volume5 pl
SourceESI -
Linearity range0.5-20 ppb

Frequently asked questions

What is the total analysis time?

The method detects all three compounds within less than 6 minutes.

What is the sensitivity for CAP at 0.5 ppb?

The signal-to-noise ratio (S/N) is 81 for CAP at 0.5 ppb.