Agilent Determination of β2-Agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges Liquid Chromatography-Tandem Mass Spectrometry Manual

The method is used for the simultaneous determination of four β2-agonist residues of terbutaline, salbutamol, clenbuterol and formoterol in pork. The method uses Agilent SampliQ SCX solid-phase extraction (SPE) cartridges and liquid chromatography-tandem mass spectrometry (LC-MS/MS) to measure the residues of four β2-agonists in pork. The method provides a sub-ng/g level of limit of detection (LOD) for all four β2-agonists in pork. The dynamic calibration ranges for these compounds are obtained from 0.25 to 5 ng/g. The overall recoveries range from 78% to 101% with RSD values between 1.8 and 7.2%.


Agilent Determination of β2-Agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges Liquid Chromatography-Tandem Mass Spectrometry Manual - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
323 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A15475DC1F35AF102C0002816E1137E4

About this manual

This document details an analytical method developed by Agilent Technologies for the simultaneous determination of four beta-2 agonist residues (terbutaline, salbutamol, clenbuterol, and formoterol) in pork. The study validates a procedure that combines liquid-liquid extraction with solid-phase extraction using Agilent SampliQ SCX cartridges to purify samples before quantification via Liquid Chromatography-Tandem Mass Spectrometry (LC-ESI-MS/MS).

The application note provides specific experimental conditions, including reagent preparation, sample hydrolysis protocols, and instrument parameters for the Agilent 1200 HPLC system and Agilent 6460 Triple Quadrupole LC-MS/MS. It reports validation results such as limits of detection (LOD), linearity, recovery rates ranging from 78% to 101%, and reproducibility data for the target compounds in pork matrices.

What's inside

  • Abstract
  • Introduction
  • Reagents and Chemicals
  • Equipment and Materials
  • Sample Preparation
  • Solid-Phase Extraction
  • Instrument Conditions
  • Results and Discussion
  • Conclusions

Specifications

Cartridge Part Number5982-3236
Column Part Number959741-906
Manifold Part Number5982-9120
Flow Rate0.4 mL/min
Column Temperature40°C

Frequently asked questions

What beta-agonists are determined in this method?

The method determines terbutaline, salbutamol, clenbuterol, and formoterol.

What is the limit of detection (LOD) for clenbuterol?

The LOD for clenbuterol is 0.02 ng/g.

What is the recovery range for the analytes?

The overall recoveries range from 78 to 101%.