Agilent Isocratic Stevia Sweetener Analysis Application Note

Two isocratic reversed-phase methods that have distinct selectivity, and use popular UV/MS or UV detection, are presented for analyzing Rebaudioside A and other steviol glycosides. A Rapid Resolution High Throughput (RRHT) Eclipse Plus Phenyl-Hexyl method using UV and MS detection resolved many compounds from a Stevia rebaudi- ana Bertoni plant extract instead of the other approach employing an amino column with a refractive index (RI) or other special detector. The RRHT Eclipse Plus Phenyl- Hexyl method’s advantages include high peak capacity, isocratic mobile phase, low acetonitrile consumption, and unique selectivity compared to a ZORBAX Carbohydrate Analysis column method. Three commercially available Stevia sweeteners were ana- lyzed for Rebaudioside A and Stevioside, the two major diterpenoid glycosides present in stevia leaves.


Agilent Isocratic Stevia Sweetener Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
359 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
AB334FC5E9E36CDC5D90B82D74C39237

About this manual

The document describes two isocratic reversed-phase liquid chromatography methods for analyzing Rebaudioside A and other steviol glycosides in Stevia sweeteners. The primary method utilizes an Agilent ZORBAX Eclipse Plus Phenyl-Hexyl column coupled with UV/MS detection, offering high peak capacity and low acetonitrile consumption. An alternative method using a ZORBAX Carbohydrate Analysis column is also presented as a secondary confirmatory technique due to its distinct selectivity.

What's inside

  • Introduction
  • Samples
  • ZORBAX Eclipse Plus Phenyl-Hexyl method
  • ZORBAX Carbohydrate Analysis column method
  • Results and Discussion
  • Conclusion

Specifications

Mobile Phase (Primary)Water (0.1% formic acid): ACN (0.1% formic acid) (82:18)
Flow Rate (Primary)0.42 mL/min
Detection Wavelength (Primary)204 nm
Temperature (Primary)35 °C
Mobile Phase (Secondary)Water: ACN (22:78)
Flow Rate (Secondary)1.5 mL/min
Temperature (Secondary)30 °C

Frequently asked questions

What detectors are used in the primary method?

The primary method uses UV/MS or UV detection, specifically a DAD and a single quad MSD with electrospray ionization.

Why is the ZORBAX Carbohydrate Analysis column recommended as a secondary method?

It is recommended as a secondary or confirmatory method due to its significantly different selectivity compared to the Eclipse Plus Phenyl-Hexyl column.

What are the advantages of the isocratic mobile phase?

The isocratic mobile phase offers no reequilibration time and eliminates baseline drift common with gradients.