Agilent 6410 Determination of Tetracyclines in Milk Application Note

A high performance liquid chromatography tandem mass spectrometric (HPLC/MS/MS) method was developed for the simultaneous determination of 10 antibiotic residues: minocycline, 4-epioxytetracycline, 4-epitetracycline, tetracycline, 4-epichlortetracycline, demeclocycline, chlortetracycline, methacycline, doxycycline, oxytetracycline in milk and animal tissues. In the method, Agilent’s novel solid phase extraction cartridge and a reversed phase Agilent ZORBAX RX C8 column (5 µm, 150 mm × 2.1 mm) were used for purification and separation. The limit of detection (LOD) is between 0.5 and 10.0 µg/kg and the limit of quantitation (LOQ) is less than 50 µg/kg. The linearity is obtained from 5 to 1000 µg/kg. Overall recoveries are between 76.4% and 101% with a relative standard deviation (RSD, n = 6) less than 8.4%. The method is rapid, sensitive, convenient and robust, and can be used to simultaneously confirm multi-residues of tetracyclines and their metabolites in milk.


Agilent 6410 Determination of Tetracyclines in Milk Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
501 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
912CE9D0293C205F788EBE0BF793045F

About this manual

This document details a High Performance Liquid Chromatography-Tandem Mass Spectrometry (HPLC/MS/MS) method developed to simultaneously determine ten antibiotic residues, specifically tetracyclines and their metabolites, in milk and animal tissues. The study utilizes the Agilent 6410 LC/MS/MS system along with a ZORBAX RX C8 column and SampliQ OPT solid phase extraction cartridges.

What's inside

  • Introduction
  • Experimental Reagents
  • Instrument Settings
  • Sample Preparation
  • Optimization of Fragmentor Voltage
  • Collision Energy Optimization
  • Optimizer Program Usage
  • Separation Challenges

Frequently asked questions

What is the limit of detection (LOD) for this method?

The limit of detection (LOD) is between 0.5 and 10.0 yg/ kg.

What is the total run time for the analysis?

The total run time is 28 minutes.

Which column is used for separation in this method?

A reversed phase Agilent ZORBAX RX C8 column (5 um, 150 mm x 2.1 mm) is used.