Agilent 6410 LC/MS Application Note: Determining Malachite Green and Leucomalachite Green in Food

This application note describes a complete method to rapidly and precisely determine residue levels of malachite green and leucomalachite green in fish with the new Agilent 6410 LC/MS triple quadrupole system. Using positive mode electrospray ionization (ESI+) and multiple reaction monitoring (MRM), qualification and quantification were accomplished without the traditional tedious PbO2 oxidation process. The LC/MS/MS method’s LOQ is 0.01 µg/Kg, which easily meets the import requirement of 2 µg/Kg set by Japan and the EU.


Agilent 6410 LC/MS Application Note: Determining Malachite Green and Leucomalachite Green in Food - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
247 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
80F12AD743CAA7844E97189A3309196F

About this manual

This document presents a complete method for rapidly and precisely determining residue levels of malachite green (MG) and leucomalachite green (LMG) in fish tissue. The method utilizes the Agilent 6410 LC/MS triple quadrupole system with positive mode electrospray ionization (ESI+) and multiple reaction monitoring (MRM). It eliminates the need for traditional post-column oxidation processes.

The application details the sample preparation procedure, including extraction with acetonitrile and cleanup with methylene chloride. It specifies the LC/MS/MS instrumental parameters, such as column dimensions, mobile phase composition, gradient elution, and specific MRM transition parameters for both compounds. The document reports a limit of quantification (LOQ) of 0.01 µg/Kg, which meets import requirements set by Japan and the EU.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Calibration Solutions
  • Sample Preparation
  • Instrumentation
  • MRM Method Parameters
  • Results and Discussion
  • Conclusions
  • References

Specifications

Limit of Quantification (LOQ)0.01 µg/Kg
Column Dimensions2.1 x 150 mm, 5 µm
Column Temperature40°C
Mobile Phase A10 mmol/L ammonium acetate (pH 4.5)
Mobile Phase BAcetonitrile
Column Flow Rate0.3 mL/min
Injection Volume10 µL
Ionization ModeESI(+)
Capillary Voltage4000 V
Drying Gas Flow11 L/min
Gas Temperature350 °C
Collision GasNitrogen

Frequently asked questions

What is the limit of quantification for this method?

The method's limit of quantification is 0.01 µg/Kg.

Does this method require post-column oxidation?

No, qualification and quantification are accomplished without the traditional tedious post-column oxidation process.

Which instrument is used for this analysis?

The analysis is performed using the Agilent 6410 LC/MS triple quadrupole system.