Agilent 7500cx ICP-MS Measurement of Trace Elements in Malt Spirit Beverages

A method for the measurement of trace elements in malt spirits (whisky) is described with reference to six different samples.


Agilent 7500cx ICP-MS Measurement of Trace Elements in Malt Spirit Beverages - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
431 KB
Published
12 June, 2012
Updated
30 September, 2023
Models
7500CX, 7500c
MD5 checksum
5183C0483290C7837C3B38945627A873

About this manual

The document details a method for measuring trace elements in malt spirits, specifically whisky, using an Agilent 7500cx ICP-MS equipped with the Octopole Reaction System (ORS). The study analyzed six different whisky samples to demonstrate the instrument's ability to remove interferences and achieve excellent spike recoveries (97-107%) and precision (<2%).

The analysis involved sample preparation using a 5x dilution with nitric and hydrochloric acid, followed by instrumental analysis using helium, hydrogen, and no-gas modes. The ORS in helium mode was found to effectively remove carbon-based and chloride-based interferences, such as those affecting Chromium and Vanadium. Hydrogen mode was utilized for Selenium analysis to improve detection limits at ultratrace levels. The study confirms the suitability of the 7500cx for routine quality control of trace metals in alcoholic beverages.

What's inside

  • Abstract
  • Introduction
  • Application
  • Experimental
  • Sample Preparation
  • Instrumental Conditions
  • Data Acquisition
  • Results and Discussion
  • Detection Limits
  • Calibration Profiles

Specifications

RF power1550 W
Sampling depth8mm
Carrier gas flow0.68 L/min
Makeup gas flow0.33 L/min
Spray chamber temperature15°C
Helium cell gas flow5.5 mL/min
Hydrogen cell gas flow40 mL/min

Frequently asked questions

What system is used to remove interferences in the 7500cx?

The Octopole Reaction System (ORS) collision/reaction cell is used to remove matrix-based polyatomic interferences.

What is the sample uptake rate for the analysis?

The sample uptake rate is approximately 150 L/min.

Which cell mode is recommended for Selenium measurement?

Hydrogen mode is recommended for Selenium to improve detection limits at low ppt levels.

Download PDF (431 KB)