This article describes the use of RTL and a three-way splitter to identify unknowns in strawberry extract.
The document details a method for analyzing complex fruit and vegetable extracts, specifically strawberry extract. It utilizes a GC/MS system equipped with a three-way splitter that directs column effluent to two GC detectors (ECD and DFPD) and an MSD simultaneously.
A specific unknown peak was identified in the sample by combining retention time locking with element-selective detector chromatograms and searching the RTL pesticide database, successfully identifying the compound as temephos without relying on a full-scan mass spectrum.
The analysis uses a 0.5-minute window centered at 41 minutes, which covers the range from 40.5 to 41.5 minutes.
The unknown peak was confirmed to be temephos, which has a target retention time of 40.74 minutes.
The confirmation uses target ion 466 and three qualifier ions with values of 125, 93, and 109.