Agilent 1200 Series Rapid Resolution LC Herbicide Analysis Application Note

This application describes how to use automated online trace enrichment (SPE) cartridges and a rapid switching valve in a Rapid Resolution LC (RRLC) system for rapid analysis of chlorinated phenoxy acid herbicides. Using small 3.5- and 1.8-µm ZORBAX RRHT columns and an automated online SPE cartridge in place of the loading pump specified in the EPA method dramatically reduces the total analysis time and virtually eliminates the potential for sample cross-contamination.


Agilent 1200 Series Rapid Resolution LC Herbicide Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
138 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
256C8C1A2EFA66624BA8F1CB03D98B19

About this manual

This document details a method for the rapid analysis of chlorinated phenoxy acid herbicides in drinking water using EPA Method 555. It focuses on the use of small ZORBAX columns (3.5 and 1.8 µm) combined with an autoSPE (Solid Phase Extraction) cartridge and automated switching valve to reduce analysis time and eliminate sample cross-contamination.

The text provides experimental conditions for the Agilent 1200 Series Rapid Resolution LC system, including specific pumps, autosamplers, and detectors. It compares standard separation methods with trace enrichment procedures, demonstrating how scaling to smaller columns reduces solvent consumption and analysis time from 60 minutes to as low as 36 minutes.

What's inside

  • Abstract
  • Introduction
  • Experimental Conditions
  • Results
  • Trace enrichment autoSPE scheme
  • Gradient separation
  • Trace enrichment chromatograms

Specifications

SystemAgilent 1200 Series Rapid Resolution LC
SoftwareChemStation 32-bit version B.02.01
Organic SolventAcetonitrile
DetectionUV 230 nm
Sample Concentration100 pg/mL in methanol or 20 ng/L (20 ppb) in reagent water

Frequently asked questions

What herbicides are analyzed in this method?

The method analyzes EPA 555 Group A chlorinated phenoxy acid herbicides, including picloram, chloramben, dicamba, bentazon, 2,4-D, dichlorprop, 2,4,5-TP, and acifluorfen.

How does the method reduce analysis time?

The method uses small ZORBAX columns (3.5 and 1.8 µm) and online trace enrichment, reducing the total analysis time from 60 minutes to 36 minutes.

What is the benefit of using online SPE?

Online SPE assures 100% sample transfer to the analysis column and dramatically increases sensitivity by increasing the analyte mass delivered to the column.