Agilent 1100 Series LC/MS Hydroxymethylfurfural Determination Application Note

This paper introduces a rapid LC/MS method for determining hydroxymethylfurfural in foods. The method has high sensitivity, selectivity, and a low detection limit


Agilent 1100 Series LC/MS Hydroxymethylfurfural Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
143 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
C9F325CC49108B450A77E5C4CDFA9175

About this manual

This document presents a sensitive and selective liquid chromatography-mass spectrometry (LC/MS) method for monitoring hydroxymethylfurfural (HMF) in food products. HMF is an indicator of food deterioration, and this method addresses toxicological concerns by providing a reliable way to detect it at low concentrations.

The procedure involves aqueous extraction of the sample, cleanup using solid-phase extraction (SPE) with an Oasis HLB cartridge, and analysis using an Agilent 1100 series HPLC system coupled with an Agilent 1100 MS detector. The method utilizes positive APCI ionization and selected ion monitoring (SIM) to achieve a detection limit of 0.005 µg/g.

What's inside

  • Introduction
  • Experimental
  • MS conditions
  • Sample Preparation
  • Results and Discussion
  • Conclusion

Specifications

ColumnZORBAX Bonus RP Narrow Bore
Flow Rate0.2 mL/min
Ionization ModePositive APCI
Mobile Phase0.01 mM acetic acid in 0.2% aqueous solution of formic acid

Frequently asked questions

Which column is used for the separation?

The separation is performed on a ZORBAX Bonus RP Narrow Bore column.

What is the flow rate used in the analysis?

The flow rate is set to 0.2 mL/min.