Agilent Improving the Effectiveness of Method Translation for Fast High Resolution Separations

The increased availability of sub-2-micron (STM) columns and increased demand for methods friendly to mass spectrometers has led to a strong trend toward conversion of existing HPLC methods to smaller diameter and smaller particle size columns. While the conversion is a simple mathematical exercise requiring the scaling flow rates, gradient times, and injection volumes, many users observe less than perfect results. Here we look closely at the problem and propose calculations that improve the speed and/or resolution in a more predictable and beneficial way.


Agilent Improving the Effectiveness of Method Translation for Fast High Resolution Separations - cover page
Brand
Agilent
Category
Other
Document type
Other
Language
English
Pages
6
File format
PDF
File size
168 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
6C1EFBA5B37A2C41C8D0FD73F14D655F

About this manual

This document is a technical application note authored by Michael Woodman of Agilent Technologies. It addresses the challenges of converting existing High-Performance Liquid Chromatography (HPLC) methods to modern sub-2-micron columns, which offer reduced analysis time and better compatibility with mass spectrometry.

The text details mathematical approaches to scaling flow rates, gradient times, and injection volumes. It specifically discusses the relationship between column efficiency and linear velocity using van Deemter plots, and highlights the impact of extracolumn dispersion on small particle size columns. The document includes experimental results demonstrating these principles using Agilent 1200 Series Rapid Resolution LC systems and ZORBAX SB-C18 columns.

What's inside

  • Method Translation
  • Sub-2-micron Columns
  • Van Deemter Plots
  • Extracolumn Dispersion
  • Gradient Slope
  • LC/MS Compatibility
  • Agilent 1200 Series
  • ZORBAX SB-C18 Columns
  • EPA Method 555
  • Solvent Consumption

Specifications

SystemAgilent 1200 Series Rapid Resolution LC
SoftwareChemStation 32-bit version B.02.01
Column 2Agilent ZORBAX SB-C18, 3.0 mm x 150 mm, 3.5 um
Aqueous Solvent25 mM phosphoric acid in Milli-Q water
Organic SolventAcetonitrile
SampleStandard mixture of chlorinated phenoxy acid herbicides
Column Temperature25 °C

Frequently asked questions

What are the benefits of converting HPLC methods to sub-2-micron columns?

Benefits include reduced analysis time, reduced solvent consumption, improved sensitivity, and greater compatibility with mass spectrometer ionization sources.

What is the recommended way to express gradient slope for method conversion?

It is more useful to express gradient slope as % change per column volume rather than % change per minute to accurately render new gradient conditions.