Agilent 1100 Series LC/MSD TOF Application Note for Sudan Red Screening

This application note presents the capability of the Agilent Time of Flight Mass Spectrometer (LC/MSD TOF) to be used as a screening, confirmation and quantitation tool-within-one analytical run of 5 min.


Agilent 1100 Series LC/MSD TOF Application Note for Sudan Red Screening - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
239 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
B2EA6A0F80F44B2D00A305419207B937

About this manual

This document presents the capability of the Agilent Time of Flight Mass Spectrometer (LC/MSD TOF) to screen for and quantify Sudan Red I, II, III, and IV in food matrices within a 5-minute analytical run. The text details the regulatory background regarding the ban on these carcinogenic dyes and the necessity for rapid detection in the food industry.

The note provides specific experimental conditions for the Agilent 1100 Series LC/MSD TOF system, including LC parameters using a ZORBAX XDB C18 column and MS settings with ESI positive ionization. It demonstrates how accurate mass determination (<3 ppm) and extracted ion chromatogram (EIC) techniques enhance the confidence in identifying these banned dyes in complex samples like chili powder and spice oil.

What's inside

  • Application
  • Introduction
  • Experimental
  • Results
  • Reference Mass Introduction
  • Empirical Formula Calculator
  • Extracted Ion Chromatogram
  • In-source Collision-Induced Dissociation

Specifications

Mass Accuracy<3 ppm
Analysis Time5 min
ColumnZORBAX XDB C18, 2.1 mm x 50 mm, 1.8 um
Flow Rate0.5 mL/min
Gas Temp350 °C
Drying Gas8 L/min
Nebulizer Pressure46 psi
Capillary Voltage4000V

Frequently asked questions

What is the analysis time for Sudan Red screening?

The analytical run for screening, confirmation, and quantitation of Sudan Reds is 5 minutes.

Which Sudan Red dyes are covered in this note?

The document covers Sudan Red I, II, III, and IV.

What is the mass accuracy of the Agilent TOF MS?

The TOF mass spectrometer provides accurate mass determination with less than 3 ppm error.