This application note from Agilent Technologies describes a simple and fast method for monitoring nitrobenzene in water using the transportable Agilent 5975T LTM GC/MSD. Liquid solvent extraction is used to prepare the sample, and ppb (ng/mL) levels of nitrobenzene can be detected in less than 5 minutes, with a calculated method detection limit of 0.017 µg/mL for standard solution.
The method uses a narrow bore (0.18 mm id) LTM column, HP-5ms 8.5 m x 0.18 mm, 0.18 µm, which reduces the run time to 4 minutes and gives injection-to-injection cycle times of around 5 minutes. The document covers the GC/MS instrument conditions, calibration from 0.1 to 10 µg/mL, method detection limit, repeatability, sample preparation and recovery, and the formulas used to calculate recovery and nitrobenzene concentration in water.
| Injector temperature | 200 °C |
|---|---|
| Column flow | 1 mL/min (He) |
| Oven temperature | 250 °C |
| Split ratio | 30:1 |
| LTM column | HP-5ms 8.5 m x 0.18 mm, 0.18 µm |
| Run time | 4 min |
| Injection-to-injection cycle time | around 5 min |
| Method detection limit | 0.017 µg/mL for standard solution |
| Calibration range | 0.1 to 10 µg/mL |
| Recovery | 108% |
The method detection limit of nitrobenzene is 0.017 µg/mL for standard solution, established from seven repeated measurements multiplied by the confidence coefficient 3.14.
The narrow bore LTM column reduces the run time to 4 minutes, and injection-to-injection cycle times are around 5 minutes.
Fifty milliliters of water sample are placed in a separating funnel with 2 mL of methylene chloride, agitated manually for one minute, and the methylene chloride layer is collected.
A 2 mL amount of 2.5 µg/mL standard solution was spiked into 50 mL water, and the calculated recovery was 108%.