Agilent 1290 Infinity LC System Application Note

This Application Note describes the separation of a trypsin-digested protein sample, using the Agilent 1290 Infinity LC System with Agilent ZORBAX Eclipse Plus C18, 150 mm × 2.1 mm, 1.8 µm particles column. Peak capacity variation with change in flow rate and time for various gradients were evaluated. Increase in flow rate from 0.2mL/min to 1.2mL/min effectively improved the peak capacity value to double and increase in gradient time from 5minutes to 50 minutes enhanced a 3.4 fold improve- ment in peak capacity value. The unique design of the Agilent 1290 Infinity LC System tolerates 1200 bar column back pressure, allowing higher flow rates, and reducing the run time.


Agilent 1290 Infinity LC System Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
780 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
54668DE3FCAE6EE7AC2F1B105002F8BB

About this manual

This document describes the separation of a trypsin-digested protein sample using the Agilent 1290 Infinity LC System with Agilent ZORBAX Eclipse Plus C18 columns. The study evaluates how peak capacity varies with changes in flow rate and gradient time.

The results indicate that increasing the flow rate from 0.2 mL/min to 1.2 mL/min effectively doubles the peak capacity value. Additionally, increasing the gradient time from 5 minutes to 50 minutes enhances the peak capacity value by 3.4-fold. The unique design of the Agilent 1290 Infinity LC System tolerates 1200 bar column back pressure, allowing for higher flow rates and reduced run times.

What's inside

  • Abstract
  • Introduction
  • System Configuration
  • LC Method parameters
  • Results and discussion
  • Conclusions

Specifications

ColumnAgilent ZORBAX Eclipse Plus C18, 150 mm x 2.1 mm, 1.8 pm
Column oven temperature50 °C
Detection wavelength214/4 nm
Reference wavelength400/60 nm
Peak width response time0.062 s
Data acquisition rate80 Hz

Frequently asked questions

How does increasing flow rate affect peak capacity?

Increasing the flow rate from 0.2 mL/min to 1.2 mL/min effectively improves the peak capacity value to double.

How does increasing gradient time affect peak capacity?

Increasing the gradient time from 5 minutes to 50 minutes enhances the peak capacity value by 3.4-fold.

Download PDF (780 KB)