Agilent Omega Platform Tube Application Note for Lead Determination in Blood

This application note describes a graphite tube design that addresses these limitations--with improved sensitivities and detection limits compared to previous tube designs


Agilent Omega Platform Tube Application Note for Lead Determination in Blood - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
410 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
19D39BB403B501FDD04F557632B3C075

About this manual

The document describes a graphite tube design called the Agilent Omega Platform Tube, which addresses limitations of previous designs such as small sample volume and restricted atomization temperature limits. This design features an L'vov platform to delay atomization until the gas environment is stable, preventing condensation and recombination of analytes.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • Results and Discussion

Frequently asked questions

What blood lead level is considered a health concern by the CDC?

The Centers for Disease Control and Prevention state that a blood lead level of greater than 10 g/dL (equivalent to 0.48 pmol/L) is of concern.

What wavelength was used for the lead measurement in this application?

The instrument settings specify a wavelength of 283.3 nm for the optical lamp.

What is the total volume capacity of the sample dispenser used?

The Agilent PSD 120 Programmable Sample Dispenser provides the capacity for up to 135 solutions.