Varian 1200L Triple Quadrupole Mass Spectrometer Perchlorate Analysis Application Note

This application note describes a method for the analysis of perchlorate using the Varian 1200L triple quadrupole mass spectrometer. The method has high selectivity and sensitivity and can be used to analyze perchlorate in contaminated water sources.


Varian 1200L Triple Quadrupole Mass Spectrometer Perchlorate Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
225 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
1200L
MD5 checksum
C4EB0976347A1C33D42B5BD348606C6E

About this manual

The document presents a method for analyzing perchlorate, a water-soluble anion found in munitions, rocket propellants, and fertilizers, which poses environmental and health risks. The method utilizes the Varian 1200L triple quadrupole mass spectrometer equipped with an ESI source and Varian ProStar 210 binary solvent delivery modules. The analysis involves filtering and treating water and produce samples to remove free chlorine and sulfate before homogenization and sonication.

Experimental conditions include the use of a Dionex AS-16 column with a mobile phase of 65mM KOH and Methanol. The method operates in negative ionization mode with Multiple Reaction Monitoring (MRM) transitions monitored simultaneously for quantitation. Results demonstrate a linear calibration curve from 0.2 to 5 ppb and a detection limit of 14 ppt for perchlorate in various matrices, including tap water, celery, zucchini, green beans, and date sugar.

What's inside

  • Introduction
  • Instrumentation
  • Experimental
  • HPLC Conditions
  • MS Parameters
  • Results
  • Conclusion

Specifications

Mobile Phase A65mM KOH
Mobile Phase BMethanol
Flow Rate300 pL/min
Mobile Phase Ratio60:40 A:B
Suppressor Regenerant72mM HpSO4
Calibration Range0.2 to 5 ppb
Correlation Coefficientr=0.9998
Detection Limit14 ppt

Frequently asked questions

What is the detection limit for perchlorate analysis?

The detection limit for perchlorate analysis was calculated to be 14 ppt.

What is the linear range for perchlorate quantitation?

The calibration curve for quantitation of perchlorate is linear from 0.2 to 5 ppb.