Agilent 320-MS Triple Quadrupole LC/MS/MS Application Note for Phenoxy Herbicides

This application note describes a method for the direct aqueous injection and automated on-line solid phase extraction (SPE) analysis of phenoxy herbicides in drinking water using the Varian 320-MS triple quadrupole LC/MS/MS. The method provides high sensitivity, selectivity, and analysis speed, meeting the requirements for trace-level quantitation of phenoxy herbicides in drinking water.


Agilent 320-MS Triple Quadrupole LC/MS/MS Application Note for Phenoxy Herbicides - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
111 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
320-MS
MD5 checksum
C5FB810B8B20565DAEA60E72F11E0557

About this manual

This document describes an analytical method for the trace-level quantitation of 17 phenoxy herbicides in drinking water. The method utilizes a fully automated, on-line Solid Phase Extraction (SPE) system coupled with the Varian (now Agilent) 320-MS Triple Quadrupole LC/MS/MS system. This approach allows for the analysis of the complete sample rather than an aliquot, improving detection capabilities while saving time and labor. The system prevents contamination and evaporation losses due to its closed nature.

The method achieves a limit of detection (LOD) of less than 20 ng/L for all studied compounds. Reproducibility over 10 injections at a routine surveillance level of 100 ng/L is less than 11% RSD. The analysis uses a 26-minute run time with linear calibration curves ranging from 20 to 500 ng/L.

What's inside

  • Introduction
  • Instrumentation
  • Sample Preparation
  • HPLC Conditions
  • Programs
  • MS Parameters
  • Results and Discussion
  • Conclusion

Specifications

Limit of Detection (LOD)less than 20 ng/L
Calibration Range20 to 500 ng/L
Reproducibility (RSD)less than 11%
Run Time26 minutes
Injection Volume10 mL
Analytical ColumnPolaris C18, 3 um, 100 x 2.0 mm

Frequently asked questions

What is the limit of detection for phenoxy herbicides using this method?

The limit of detection is less than 20 ng/L for all compounds studied.

How many replicate injections were performed to test reproducibility?

Ten replicate injections were performed at the routine surveillance level of 100 ng/L.