Agilent The determination of Cd Cr Cu Ni Pb in a concentrated CaCl2/NaCl solution by AAS Application Note

The document describes the determination of Cd, Cr, Cu, Ni, and Pb in a concentrated CaCl2/NaCl solution using AAS. The analysis of such a high solids containing solution presents challenges, which are addressed through indirect flame atomization, direct flame atomization, and graphite furnace atomization methods. It is important to clean the nebulizer and measure a reagent blank during the analysis. The results are compared against a calibration graph. The document provides calibration data for Cd.


Agilent The determination of Cd Cr Cu Ni Pb in a concentrated CaCl2/NaCl solution by AAS Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
75 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
7162286C8596FD8051A8DE0B04278DFD

About this manual

The document addresses the analytical challenge of measuring trace heavy metals in a high-solids caustic suspension containing calcium chloride and sodium chloride. It highlights issues such as burner blockage and nebulizer damage caused by the matrix, necessitating specific measurement techniques.

Three methods are evaluated: indirect flame atomization, direct flame atomization, and graphite furnace atomization. The text provides detailed instrumentation parameters, chemical preparation steps, calibration data tables, and comparative results to demonstrate that the graphite furnace technique is best suited for achieving the required low detection limits.

What's inside

  • Summary
  • Instrumentation and Chemicals
  • Pre-concentration and Flame Measurements
  • Co-precipitation Pre-concentration of Cr
  • Direct Flame Measurements
  • Direct Graphite Furnace Measurements
  • Conclusion

Specifications

Instrument ModelAgilent SpectrAA20B Atomic Absorption Spectrometer
Graphite Furnace ModelGTA-96 graphite furnace atomizer

Frequently asked questions

Which method is best suited for determining heavy metals in this solution?

The graphite furnace technique is best suited because the sensitivity of flame AAS was often inadequate due to low detection limits.

What elements are determined in the solution?

The heavy metals Cd, Cr, Cu, Ni and Pb must be determined.

Download PDF (75 KB)

More Agilent Manuals