Agilent GTA Graphite Furnace Forked Platform Design Application Note

describes the evaluation of a new forked platform design for the Agilent GTA graphite furnaces


Agilent GTA Graphite Furnace Forked Platform Design Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
115 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
97D7C61593A7E61E232F8A4C9EE8FC1A

About this manual

This document evaluates a new forked platform design intended for use with Agilent GTA graphite furnaces. The design features a partitioned tube with notches that allow the platform to be inserted with minimal physical contact to the tube wall, preventing rotation during insertion. This design aims to reduce chemical and background interferences by delaying sample vaporization relative to the tube wall temperature.

The evaluation tested the platform's performance by analyzing five elements (As, Se, Pb, Cd, and Sn) using Agilent SpectrAA-400 spectrometers. Results indicated that the forked platform provided thermal delays similar to current platforms for most elements, with peak area sensitivities within 10% of the regular platform. Accuracy tests using an EPA standard and analysis of a difficult environmental sample showed no significant difference in results compared to the current platform design.

What's inside

  • Introduction
  • Methodology
  • Results
  • Conclusion
  • Forked Platform Design
  • Graphite Furnace Techniques
  • Trace Element Analysis
  • EPA Standard Evaluation
  • Environmental Sample Analysis
  • Instrument Operating Parameters

Specifications

Elements AnalyzedAs, Se, Pb, Cd, Sn
Instrument ModelsSpectrAA-400 Zeeman and SpectrAA-400 Deuterium
Document DateNovember 1, 2010

Frequently asked questions

What are the advantages of the forked platform design?

Advantages include simple positioning of the tube, easy alignment of the sample dispensing capillary tip, and stability of the platform during operation.

What elements were used to evaluate the forked platform?

The evaluation examined the determination of five elements: As, Se, Pb, Cd, and Sn.

What spectrometers were used for the Cd and Sn measurements?

The Zeeman spectrometer was used for the Cd and Sn measurements.

What is the atomization temperature for Lead (Pb)?

The atomization temperature for Lead is 2300 degrees.