Agilent GTA Graphite Furnace Forked Platform Application Note

The use of the forked platform graphite furnace atomic absorption spectrometry method for the determination of trace elements in stainless steel is described. This method has high sensitivity and low background interference.


Agilent GTA Graphite Furnace Forked Platform Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
106 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
239F5504A92F1AB5557766BBA7734095

About this manual

The document details the analysis of trace elements (As, Se, Pb, Cd, Sb, Bi, and Ag) in stainless steel samples using Graphite Furnace Atomic Absorption Spectroscopy (GFAAS). It highlights the use of a new forked platform design installed in partitioned style tubes to minimize physical contact and reduce chemical and background interferences, particularly for volatile elements.

The methodology covers sample preparation involving the dissolution of stainless steel shavings, the use of specific chemical modifiers (such as magnesium nitrate and sulfuric acid), and standard additions calibration. The note provides instrument operating parameters, furnace parameters, and chemical modifier volumes used to achieve high sensitivity and precision in the presence of high metal matrices.

What's inside

  • Introduction
  • Methodology
  • Instrument Operating Parameters
  • Furnace Parameters
  • Chemical Modifiers
  • Results
  • Conclusion

Specifications

Major constituents concentrationFe -75%, Cr - 13%, Ni, - 8%, MO - 2%, and Al - 1%
Sample volume range3 to 5 pL
Maximum atomization ramp rate2000 °C/sec
Atomization hold time2.7 seconds

Frequently asked questions

Which elements were determined in the stainless steel samples?

The study analyzed for seven elements: As, Se, Pb, Cd, Sb, Bi, and Ag.

What chemical modifiers were used for Arsenic (As) and Selenium (Se)?

A 1% Mg(NO3)2 modifier was used for both As and Se determinations.

How does the forked platform design help with analysis?

It provides minimal physical contact between the platform and tube wall, preventing platform rotation during insertion and ensuring stability.