Agilent 500 Ion Trap LC/MS Aflatoxin Identification Application Note

This application note describes a rapid and sensitive LC/MS/MS method for the detection and analysis of four aflatoxins (B1, B2, G1, and G2) that provides detection down to 1 ppb. The 500 Ion Trap MS/MS capabilities result in the collection of clear, baseline resolved chromatographic peaks for target compound quantitation.


Agilent 500 Ion Trap LC/MS Aflatoxin Identification Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
128 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
A697F7FE793F7B690B1E34618C3DE956

About this manual

The document presents an analytical method for the detection and analysis of four carcinogenic mycotoxins: aflatoxins B1, B2, G1, and G2. The study utilizes a rapid Liquid Chromatography-Tandem Mass Spectrometry (LC/MS/MS) approach to achieve high sensitivity with detection limits down to 1 ppb.

What's inside

  • Instrumentation
  • Materials and Reagents
  • HPLC conditions
  • MS parameters
  • Results and Discussion
  • Calibration Curve Report
  • Conclusion

Frequently asked questions

What is the detection limit for aflatoxins using this method?

The method provides detection down to 1 ppb.

How long does it take to separate and identify the four aflatoxins?

The four aflatoxins were separated and identified in less than 15 minutes.

What is the linear range for the calibration curves?

Calibration curves are linear from 1 to 50 ppb.