Agilent Solid Phase Extraction of THC THC-COOH 11-OH-THC from Whole Blood Manual

This document describes the extraction and analysis method of marijuana and its metabolites in human blood, as well as the use of related instruments and reagents.


Agilent Solid Phase Extraction of THC THC-COOH 11-OH-THC from Whole Blood Manual - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
209 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
38E9577CDB47098F8A2819F8AD162AB0

About this manual

This document details an effective Solid Phase Extraction (SPE) method coupled with Gas Chromatography/Mass Spectrometry (GC/MS) for the extraction of tetrahydrocannabinol (THC), 11-hydroxy-THC, and THC-COOH from human whole blood. The procedure is designed to detect recent drug usage and impairment.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • SPE Method
  • Derivatization
  • GC/MS Conditions

Frequently asked questions

What is the minimum centrifugation speed required for sample preparation?

The sample must be centrifuged at a minimum of 2500 rpm for 5 minutes.

What is the maximum temperature allowed when evaporating elution fractions?

Elution fractions must be evaporated under nitrogen at no higher than 40°C.