Agilent 1290 Infinity LC System Application Note for Genotoxic Impurity Determination

This Application Note describes a method to analyze traces of potential genotoxic impurities (PGIs) in pharmaceuticals using the Agilent 1290 Infinity LC system and a 60 mm Agilent Max-Light Cartridge Cell, which offers maximum sensitivity.


Agilent 1290 Infinity LC System Application Note for Genotoxic Impurity Determination - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
575 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
20F62AD3C0D79167CD5E9DAB51CAF019

About this manual

The document details an analytical method developed by researchers at the Research Institute for Chromatography and Agilent Technologies to analyze traces of potential genotoxic impurities (PGIs) in pharmaceuticals. The study utilizes a generic method to separate ten specific arylamine and aminopyridine impurities using UHPLC coupled with a Diode-Array Detector (DAD).

To achieve the required low detection limits without mass spectrometry, the method employs an Agilent Max-Light Cartridge High Sensitivity Cell featuring a 60 mm optical path length. The note provides specific experimental procedures for sample preparation and lists detailed chromatographic conditions, including column types, mobile phases, gradients, and temperature settings for both generic and dedicated methods.

What's inside

  • Introduction
  • Abstract
  • Standards Solutions
  • Sample Preparation
  • Equipment
  • Chromatographic Conditions
  • Generic Method
  • Methanol Method
  • Acetonitrile Method
  • Dedicated Method

Frequently asked questions

What is the optical path length of the Max-Light Cartridge High Sensitivity Cell used?

The cell has an optical path length of 60 mm with an internal volume of only 4 pL.

What is the detection limit achieved for potential genotoxic impurities?

Detection limits were as low as 0.2 ng/mL, which corresponds to 4 pg on-column.

What is the typical analysis time for real samples using this method?

Analysis times are approximately 5 minutes when developed for a relevant combination of PGI and API.