Agilent ProSEC 300S Columns Calibration Application Note

This document introduces the key technology in protein analysis-size exclusion chromatography (SEC), and illustrates the principle, characteristics and application of SEC with Agilent ProSEC 300S column as an example


Agilent ProSEC 300S Columns Calibration Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
576 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
8EB8809FBED7522730B3107367DDF2FE

About this manual

This document details the use of size exclusion chromatography (SEC) for protein analysis, highlighting the importance of minimizing protein interactions with column packing material. It presents an experiment using a series of globular proteins to validate the performance of the ProSEC 300S column.

What's inside

  • Introduction
  • Methods and Materials
  • Results and Discussion
  • Conclusion

Specifications

Column Dimensions300 x 7.5 mm
Eluent Composition0.3M NaCl, 50mM KH2PO4-K2HPO4, pH 6.8
Flow Rate1.0 mL/min
Injection Volume20 pL
Temperature25 °C
Detection Wavelength280 nm

Frequently asked questions

What proteins were used for calibration?

The calibration used Thyroglobulin, y-Globulin, Bovine serum albumin, Ovalbumin, Carbonic anhydrase, Myoglobulin, Cytochrome C, and Bacitracin.

What does the calibration curve show?

The calibration curve shows a linear relationship between the log of the molecular weight and the retention time.