Agilent Polysiloxane Analysis Application Note

This application note introduces the analysis of polysiloxane using GPC/SEC with RI and ELS detection on Agilent PLgel 5 µm MIXED-E column. Polysiloxanes are important commercial polymers used in applications such as lubricating oils, defoaming agents, and sealants. Analyzing these materials by gel permeation chromatography (GPC) is an important part of process and quality control procedures. Although polysiloxane is soluble in tetrahydrofuran (THF), it has the same refractive index as this solvent. Therefore, refractive index (RI) detection cannot be used with THF.


Agilent Polysiloxane Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
73 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
7E1B78ABEF5725C1163EFE9F1A4AF57E

About this manual

Polysiloxanes are important commercial polymers used in various applications such as lubricating oils and sealants. Analyzing these materials via gel permeation chromatography (GPC) is crucial for process and quality control. However, because polysiloxanes are isorefractive with tetrahydrofuran (THF), standard refractive index (RI) detection cannot be used with this solvent.

The document outlines two options for analysis: changing the solvent to use RI detection or using THF with an Agilent evaporative light scattering detector (ELSD). The ELSD provides a positive response proportional to sample concentration in all solvents and does not generate system peaks due to solvent imbalance. The note includes specific conditions for the analysis, including column type, flow rate, and detector settings, and compares the data quality between RI and ELS detection.

What's inside

  • Introduction
  • Conditions
  • Analysis of Polysiloxane
  • System peaks
  • Conclusions

Specifications

Sample concentration0.2% Sw/v
Column3 x Agilent PLgel 5 pm MIXED-E, 300 mm = 7.5 mm
EluentToluene or THF
Flow rate1.0 mL/min
Injection volume100 pL
Detector Neb temperature40 °C
Detector Evap temperature80 °C
Detector Gas flow1.0 SLM

Frequently asked questions

Why is RI detection not used with THF for polysiloxane analysis?

Polysiloxane is isorefractive with the tetrahydrofuran (THF) solvent, making RI detection ineffective.

What are the advantages of using an ELSD over RI detection?

ELSD does not rely on refractive index, works in all solvents, does not generate system peaks, and offers higher sensitivity.

How does the sensitivity of the ELSD compare to the RI detector in this application?

The ELS sensitivity is about 250 times higher than the RI detector.