Agilent PL-SAX 1000A Strong Anion-Exchange HPLC Columns Application Note

This article introduces the use of PL-SAX strong anion-exchange HPLC columns for high-resolution separations of oligonucleotides. After deprotection of the oligonucleotide product, hydrophobicity-based separation can be performed, followed by anion-exchange HPLC for further resolution. The experimental results show that the PL-SAX column can achieve separation of n-1 from n.


Agilent PL-SAX 1000A Strong Anion-Exchange HPLC Columns Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
231 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
58097F39CA4C1312862519AA3100A460

About this manual

This document describes the use of Agilent PL-SAX strong anion-exchange columns for the high-resolution separation and purification of oligonucleotides. It explains how the material's thermal and chemical stability allows for analysis under elevated temperatures and harsh conditions, such as high pH, to resolve failure sequences and modified forms like phosphorothioates.

The note provides specific experimental protocols for three different separation scenarios: separating poly-T oligonucleotides of varying lengths, purifying large synthetic oligonucleotides, and separating complete and incomplete thiolated forms. Each protocol specifies the column dimensions, eluent composition, gradient profiles, flow rates, and temperatures required to achieve baseline separation.

What's inside

  • Introduction
  • Materials and Reagents
  • Results and Discussion
  • Conclusion

Specifications

Column Part Number (50 x 4.6 mm)PL1551-1802
Column Part Number (150 x 4.6 mm)PL1551-3802
Particle Size8 ym
Flow Rate1.5 mL/min
Operating Temperature60 °C
Linear Velocity360 cm/h

Frequently asked questions

What eluent is used for separating poly-T oligonucleotides?

Eluent A is 7:93 v/v acetonitrile: 0.1 M TEAA, pH 8.5, and Eluent B is 7:93 v/v acetonitrile: 0.1 M TEAA, 1 M ammonium chloride pH 8.5.

What conditions are used to separate thiolated oligonucleotides?

The separation uses sodium hydroxide as the eluent with a gradient from 75-100% B for 25 min followed by 0-100% B in 20 min.