USP Analysis of Tetracyclines Application Note

This document describes the use of tetracyclines and the USP method1 for the analysis of the compound


USP Analysis of Tetracyclines Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
658 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
52DF40E7E745827E8EE3962563B687DA

About this manual

The document details an HPLC method for analyzing tetracycline antibiotics, specifically doxycycline, oxytetracycline, and metacycline. It explains that these compounds are analyzed using a reversed-phase PS/DVB PLRP-S column, which offers chemical stability and high pressure capability suitable for USP methods.

The note provides specific experimental conditions including sample preparation (20 mg Tetracycline in 25 mL 0.01 M HCl), column specifications (PLRP-S 100A), eluent composition, flow rate (1.0 mL/min), temperature (60°C), and UV detection at 254 nm. Results demonstrate good resolution of the tetracycline mix.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Conclusion

Frequently asked questions

What sample preparation is required for the tetracycline analysis?

The sample consists of 20 mg of tetracycline dissolved in 25 mL of 0.01 M HCl.

What are the flow rate and temperature settings for the analysis?

The analysis uses a flow rate of 1.0 mL/min at a temperature of 60°C.

At what wavelength is the detection performed?

Detection is performed using UV at 254 nm.

Download PDF (658 KB)

More Agilent Manuals