Agilent Optimized Separation of Angiotensins by HPLC with ELSD Application Note

This document describes the optimization of peptide separations using PLRP-S reversed-phase material in acidic and basic conditions by HPLC with ELSD.


Agilent Optimized Separation of Angiotensins by HPLC with ELSD Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
1.1 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
1FF3746360755EC6454CA71C1F93C727

About this manual

The document details a method for separating angiotensins, a peptide hormone mixture, utilizing PLRP-S reversed-phase material which is stable across a wide pH range (1-14). This stability allows the use of both conventional acidic ion pairing agents like trifluoroacetic acid (TFA) and alkaline conditions to optimize selectivity based on amino acid charge distribution.

The study compares separation results using different mobile phase modifiers. Under acidic conditions, varying concentrations of TFA were tested, showing that while 0.1% provides good peak shape, it causes co-elution of Angiotensin II and III. Reducing the concentration improves resolution but can cause band broadening. Alternatively, under basic conditions, ammonium hydroxide was used as a modifier to achieve separation.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Conditions
  • Results and Discussion
  • Conclusion

Frequently asked questions

What is the specified flow rate for this separation method?

The procedure uses a constant flow rate of 1.0 mL/min.

How long does the linear gradient run in Figure 2?

The separation takes 15 minutes to complete using a linear gradient from 10% to 100% B.

What are the nebulizer and evaporator temperatures for Figure 1?

The detector settings for Figure 1 require a neb temperature of 85 °C and an evap temperature of 75 °C.