Agilent Fast Reversed Phase HPLC of Analgesics Application Note

This application note describes a fast reversed-phase HPLC method for the determination of aspirin, salicylic acid and phenacetin. The method uses a highly retentive polymeric reversed-phase column (PLRP-S 100Å) and gradient elution. The method provides excellent separation of the analytes, and can be completed in less than 10 minutes. The method is also highly reproducible and stable.


Agilent Fast Reversed Phase HPLC of Analgesics Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
596 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
CD9AA3430407B836612F8E0C80B624E9

About this manual

This document presents an isocratic High-Performance Liquid Chromatography (HPLC) method designed to resolve the analgesics aspirin, salicylic acid, and phenacetin in under 10 minutes. The method addresses difficulties in quantifying these compounds when using standard UV detectors due to differing extinction coefficients.

The recommended system utilizes a PLRP-S 100A polymeric reversed phase column coupled with an Agilent evaporative light scattering detector (ELSD). This combination provides a uniform response for all analytes and allows the use of aggressive eluents across a wide pH range.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion

Frequently asked questions

What is the purpose of this HPLC method?

The method resolves aspirin, salicylic acid, and phenacetin in under 10 minutes.

What are the dimensions of the recommended column?

The PLRP-S 100A column is 150 x 4.6 mm.

What are the operating temperatures for the Agilent ELSD?

The nebulizer temperature is set to 95 °C and the evaporator to 50 °C.