Agilent Effective Analysis of Plasticizers by GPC and ELSD Application Note

An efficient plasticizer additive analysis method


Agilent Effective Analysis of Plasticizers by GPC and ELSD Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
1.3 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
F51B5AC7E0CE41BDB6E0AECA53DF988D

About this manual

This document details an application for analyzing polymers that contain plasticizers, specifically focusing on the determination of polymer molecular weight distribution and qualitative information about additives. It highlights the benefits of using Agilent Evaporative Light Scattering Detection (ELSD) over Refractive Index (RI) detection, noting that ELSD provides a significantly improved signal-to-noise ratio and is independent of dn/de effects.

The note specifies the use of PLgel 5 um MIXED-C columns for this application, citing their high efficiency and broad resolving molecular weight range. It presents results from the analysis of a polymer containing dibutylphthalate (DBP), demonstrating the system's ability to separate the polymer from the additive and a flow rate marker (toluene). The conclusion emphasizes the robust design of the Agilent ELSD and the suitability of the system for compounds with no chromophores.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Conditions
  • Results and Discussion
  • Conclusion

Specifications

Column dimensions300 x 7.5 mm
Column part numberPL1110-6500
Nebulizer temperature40 °C
Evaporator temperature90 °C
Carrier gas flow1.5 SLM
EluentTHF
Flow rate1.0 mL/min

Frequently asked questions

What is the benefit of using Agilent ELSD over RI detection for plasticizer analysis?

ELSD provides a significantly improved signal to noise ratio and the detector response is almost independent of dn/de effects.

Which columns are recommended for mid molecular weight polymers?

The PLgel 5 um MIXED-C columns are recommended due to their high efficiency and broad resolving molecular weight range.

What are the operating conditions for the Agilent ELSD in this application?

The conditions are a nebulizer temperature of 40 °C, evaporator temperature of 90 °C, and a gas flow of 1.5 SLM.