Agilent 1290 Infinity LC Application Note: Determination of Explosives

This application note describes the method for detection of nanogram levels of explosive constituents in seawater samples, after explosion of ordnance devices.


Agilent 1290 Infinity LC Application Note: Determination of Explosives - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
813 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
A2416790667DCBA4206E3F1CA39547A6

About this manual

This document details an optimized chromatographic method for detecting explosive constituents, such as TNT and hexogen, in seawater samples following ordnance detonations. The method utilizes sample enrichment on Agilent SampliQ PS-DVB cartridges followed by analysis with the Agilent 1290 Infinity LC System.

The study evaluates various RRHD- and RRHT columns to determine optimal selectivity for separating eight specific explosives and an internal standard within a short timeframe. The final optimized method achieves complete separation in approximately 7 minutes using an Agilent ZORBAX Stable Bond CN column, with detection limits improved by the use of a 60-mm cell.

What's inside

  • Introduction
  • Instrumentation Configuration
  • Sample Preparation
  • Column Screening Results
  • Chromatographic Conditions
  • Final Method Optimization

Specifications

Separation Time7.0 minutes
Retention Time Precision< 0.5% RSD
Linearity Range0.5 to 50 ng/pL
Detection Improvement Factor4x

Frequently asked questions

Which column was selected for the final optimized method?

The Agilent ZORBAX Stable Bond CN column was chosen because it provided the best separation of eight explosives and the internal standard.

What sample preparation method is used?

Samples are enriched using highly cross-linked polystyrene-divinylbenzene (PS-DVB) cartridges, such as the Agilent SampliQ.

What are the detection limits for this method?

The method allows for the detection of nanogram levels of explosive constituents in seawater samples.