Agilent Hi-Plex H Column Organic Acids in Silage Application Note

This application note describes the analysis of organic acids in silage using high-performance liquid chromatography (HPLC). The concentration of lactic acid, acetic acid, and butyric acid in silage can be determined by adding water and filtering the sample. This method is applicable for the analysis of grass, whole plant, and corn silage.


Agilent Hi-Plex H Column Organic Acids in Silage Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
290 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
413DFFE9DFCC0BB408D47656F3891331

About this manual

This document describes the analysis of organic acids in silage samples, specifically lactic, acetic, and butyric acid, using High-Performance Liquid Chromatography (HPLC). The procedure utilizes an Agilent Hi-Plex H column to separate volatile and nonvolatile acids without prior derivatization. The note covers the extraction of acids according to EN 13037 and details the chromatographic conditions used for the analysis.

What's inside

  • Introduction
  • Materials and Methods
  • Conditions
  • Results and Discussion
  • Conclusion
  • Peak identification

Specifications

Column dimensions7.7 x 300 mm
Mobile phase0.005 M H2SO4
Flow rate0.7 mL/min
Injection volume20 µL
Temperature60°C
Detector temperature55 °C
Pressure4.6 MPa

Frequently asked questions

What acids are analyzed in the silage samples?

The concentration of lactic acid, acetic acid, and butyric acid is analyzed.

What is the recommended extraction procedure for silage samples?

Add 1.25 L of water to 250 mL of silage and agitate for one hour, then filter through a 0.45 pm membrane.

What are the stated operating pressure limits for this column?

The system operates at a pressure of 4.6 MPa, which is equivalent to 46 bar or 670 psi.

What temperature settings are required for the detector and column?

The column is maintained at 60°C while the detector operates at 55°C.