About this manual

This document details an analytical method for detecting four synthetic sweeteners—acesulfame, saccharin, cyclamate, and sucralose—in waste water. The method utilizes Agilent Bond Elut Plexa solid-phase extraction (SPE) cartridges to pre-concentrate the analytes from water samples prior to detection.

What's inside

  • Introduction
  • Materials and Methods
  • SPE method
  • HPLC/MS conditions
  • Results and Discussion
  • Conclusions
  • References

Specifications

SPE Cartridge Capacity200 mg
SPE Cartridge Volume6 mL
Column Dimensions4.6 x 50 mm
Column Particle Size1.8 pm
Mobile Phase AWater, 2 mM ammonium carbonate
Mobile Phase BMethanol, 2 mM ammonium carbonate
Flow Rate0.6 mL /min
Injection Volume (20 pL)Recovery > 86%

Frequently asked questions

Which sweeteners are analyzed in this method?

The method analyzes acesulfame, saccharin, cyclamate, and sucralose.

What is the concentration of sweeteners in the water sample?

The concentration of acesulfame, cyclamate, saccharin, and sucralose in the water sample is 1 ppb.

Download PDF (246 KB)

More Agilent Manuals


View all Agilent manuals →