Agilent Hi-Plex H Column Application Note for Fermentation Analysis

This application note describes the HPLC analysis of byproducts in fermentation liquids using an Agilent Hi-Plex H column.


Agilent Hi-Plex H Column Application Note for Fermentation Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
201 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
F4737B55E43783A38395864AEF3B1B03

About this manual

This document details the use of an Agilent Hi-Plex H column for analyzing byproducts and degradation products derived from biomass fermentation, such as xylose to xylitol conversion. It presents results from two specific samples: one obtained via hydrothermal digestion of straw and another following partial evaporation and fermentation.

What's inside

  • Introduction
  • Materials and Methods
  • Conditions
  • Results
  • Conclusion

Specifications

Column Dimensions7.7 x 300 mm, 8 µm
Mobile Phase0.005 M H2SO4
Flow Rate0.7 mL/min
Injection Volume20 µL
Temperature60°C
Pressure4.6 MPa (46 bar, 670 psi)
Detector Temperature55°C

Frequently asked questions

What is the recommended mobile phase for this analysis?

The document recommends using sulfuric acid as the mobile phase.

Is the Hi-Plex H column suitable for high levels of organic acids?

Yes, the column is recommended for use with samples that contain high levels of organic acids.