Agilent 1290 Infinity LC and 6460 LC/MS Pesticide Determination Application Note

The qualitative and quantitative analysis of pesticides at trace levels in baby food matrices using UHPLC and triple quadrupole MS is demonstrated. Sample preparation is performed using an Agilent Bond Elut QuEChERS kit for extraction and dispersive SPE. The extracts are analyzed by LC/MS/MS on an Agilent 1290 Infinity LC system coupled to an Agilent 6460 triple quadrupole LC/MS using Dynamic MRM. The method and extraction performance were evaluated in terms of repeatability, linearity and sensitivity. Moreover the influence of the additional dispersive SPE cleanup was investigated. Detection limits were between 500 ng/kg and 10 ng/kg (ppt), which is much lower than the maximum residue level (MRL) of 10 µg/kg (ppb) imposed by the European Union.


Agilent 1290 Infinity LC and 6460 LC/MS Pesticide Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
817 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
74F652D7896DDFA36D3CFC44406D5285

About this manual

This document details a method for the qualitative and quantitative analysis of 40 pesticides at trace levels in baby food matrices. The method utilizes UHPLC and triple quadrupole mass spectrometry with Dynamic MRM to achieve detection limits between 500 ng/kg and 10 ng/kg.

Sample preparation involves extraction using an Agilent Bond Elut QUEChERS kit followed by dispersive SPE cleanup. The analysis is performed on an Agilent 1290 Infinity LC system coupled to an Agilent 6460 triple quadrupole LC/MS using Dynamic MRM with a total analysis time of 10 minutes.

What's inside

  • Introduction
  • Instrumentation
  • Method parameters
  • Dynamic MRM data acquisition parameters
  • Solutions and Samples
  • Sample Preparation
  • Extraction
  • Dispersive SPE
  • Results and discussion

Specifications

Total analysis time10 min
Detection limits500 ng/kg and 10 ng/kg
ColumnAgilent ZORBAX Eclipse Plus RRHD C18
Ionization modeElectrospray, positive ionization
Drying gas temperature250 °C
Capillary voltage4500 V
Cycle time200 ms

Frequently asked questions

What is the total analysis time for the pesticide determination method?

The total analysis time is 10 minutes, including re-equilibration.

Which extraction kit is used for sample preparation?

An Agilent Bond Elut QUEChERS kit is used for extraction and dispersive SPE.

What is the range of detection limits achieved?

Detection limits were between 500 ng/kg and 10 ng/kg.

How many pesticides are analyzed in this method?

The method analyzes 40 pesticides in baby food.

Download PDF (817 KB)

More Agilent Manuals


View all Agilent manuals →