Agilent ZORBAX RRHD 300SB-C8 Column Application Note

This application note describes a method for achieving highly resolved ultrafast separations of reduced and alkylated monoclonal antibodies using an Agilent ZORBAX sub-2 µm Rapid Resolution High Definition (RRHD) 300SB-C8 reversed-phase column under optimized chromatographic conditions. The StableBond C8 phase and RRHD column technology, in combination with robust gradient elution conditions, enabled efficient, fast, and effective separation of light chain and two heavy chain monoclonal antibody variants in under 4 minutes. Additionally, alternate mobile phase compositions were employed, making separations flexible for ultra fast LC/MS analysis. At elevated operating pressures and temperature, the stable C8 phase and rugged column packing technology delivered highly reproducible separations and delivered excellent chromatographic run-to-run results.


Agilent ZORBAX RRHD 300SB-C8 Column Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
664 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
5E5ED92F2C5A60DE4C003254151B3EF2

About this manual

This document describes the use of an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 reversed-phase column for the ultra-fast separation of reduced and alkylated monoclonal antibodies. The study highlights how the StableBond C8 phase and RRHD technology enable efficient separations in under four minutes at elevated temperatures.

The note details specific experimental procedures, including the reduction and alkylation of antibodies, and provides optimized UHPLC conditions using an Agilent 1290 LC Infinity system. It also explores mobile phase compositions suitable for LC/MS analysis, demonstrating flexibility with different ion pairing additives like formic acid and acetic acid.

What's inside

  • Abstract
  • Introduction
  • Materials
  • Reduction and Alkylation
  • UHPLC Conditions
  • Results and Discussion
  • Mobile phase compositions for fast enhanced LC/MS analysis

Specifications

Column Dimensions2.1 x 100 mm
Particle Size1.8 ym
Operating Temperature75 °C
Flow Rate0.5 mL/min
Detection Wavelength225 nm

Frequently asked questions

What operating temperature was used for the analysis?

The analysis was performed at an elevated temperature of 75 °C.

Which instrument system was used for the UHPLC analysis?

An Agilent 1290 LC Infinity system with auto injector, binary pump, thermostatted oven, and diode array detector was used.