Agilent SAMHSA-Compliant LC/MS/MS Analysis of 6-Acetylmorphine in Urine with Agilent Bond Elut Plexa PCX Agilent Poroshell 120 Manual

This article presents a SAMHSA-compliant LC/MS/MS analysis method for the detection of 6-acetylmorphine in urine. The method utilizes Agilent products such as Agilent Bond Elut Plexa PCX and Agilent Poroshell 120, and demonstrates simplification, linearity, limit of detection (LOD), accuracy, and precision.


Agilent SAMHSA-Compliant LC/MS/MS Analysis of 6-Acetylmorphine in Urine with Agilent Bond Elut Plexa PCX Agilent Poroshell 120 Manual - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
115 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
6D0CE3F9B6B59BB70DD31C30292BF6C8

About this manual

The document presents a simplified liquid chromatography-tandem mass spectrometry (LC/MS/MS) method for the confirmation of 6-acetylmorphine in urine, meeting US Substance Abuse and Mental Health Services Administration (SAMHSA) guidelines effective October 2010. The method utilizes Agilent Bond Elut Plexa PCX mixed-mode polymeric SPE sorbent to minimize ion suppression and maximize sensitivity.

The analytical procedure employs an Agilent Poroshell 120 EC-C18 superficially porous column for high-capacity separation and an Agilent 6460 Triple Quadrupole LC/MS system with AJST electrospray source. The method demonstrates excellent linearity, limit of detection (LOD), accuracy, precision, and extraction recovery suitable for forensic toxicology applications.

What's inside

  • Abstract
  • Introduction
  • Analytes
  • Materials and instrumentation
  • Sample preparation
  • LC/MS/MS conditions
  • Results and Discussion
  • Method evaluation

Frequently asked questions

What is the SAMHSA confirmation cutoff concentration for 6-AM?

The updated SAMHSA confirmation cutoff concentration for 6-AM is 10 ng/mL.

What is the limit of detection (LOD) required by this method?

A LOD at 10% of the cutoff would be 1 ng/mL.

What is the extraction recovery for this method?

The extraction procedure yields an extraction recovery of 83%.