Agilent Low-level Speciated Analysis of Cr(III) and Cr(VI) Using LC(IC)-ICP-MS Application Note

This application note describes the use of LC(IC)-ICP-MS for low-level speciation analysis of Cr(III) and Cr(VI).


Agilent Low-level Speciated Analysis of Cr(III) and Cr(VI) Using LC(IC)-ICP-MS Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
259 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
C7630A49FE3E412E8B18B1E9CAB0AA1D

About this manual

Chromium exists primarily as trivalent Cr(III) or hexavalent Cr(VI), which have different toxicological properties. This document details a routine, highly sensitive method to determine both species in various sample types using anion exchange liquid chromatography (LC/IC) coupled with Inductively Coupled Plasma Mass Spectrometry (ICP-MS). The method utilizes the Agilent 7700x ICP-MS equipped with the Octopole Reaction System (ORS) operated in helium mode to remove matrix-based interferences.

The study demonstrates the analysis of mineral waters and California tap water. It reports detection limits below 200 ng/L for both species under standard conditions, and down to 0.008 pg/L for Cr(VI) alone using an optimized method with larger injection volumes. The document provides specific operating parameters for the HPLC and ICP-MS instruments used in the experiments.

What's inside

  • Introduction
  • Experimental
  • HPLC conditions
  • ICP-MS conditions
  • Results and discussion
  • Drinking water analysis
  • Quantification of Cr(VI) at ultra-trace levels

Specifications

Column dimensions4.6 mm x 30 mm
Flow rate1.2 mL/min
Injection volume (standard)100 µL
RF power1550 W
Carrier gas flow rate1.05 L/min
Cell gas type and flowHe at 4 mL/min
Detection limit (Cr(VI) optimized)~0.008 pg/L

Frequently asked questions

How does the method remove matrix interferences?

The ORS collision/reaction cell is operated in helium mode to remove matrix-based polyatomic interferences such as ArC and ClO.

What is the detection limit for Cr(VI) using the modified method?

The detection limit for Cr(VI) was reduced to 0.008 pg/L.

How long does the analysis take using anion exchange chromatography?

The analysis is rapid, taking only about 3 minutes.