This document introduces the Agilent developed methods for the rapid, sensitive and robust detection of phthalates in food.
This document outlines highly sensitive and reliable methods developed by Agilent Technologies for the detection of phthalate contaminants in food. These methods are designed to run on a wide range of mass spectrometry platforms, including both single and triple quadrupole instruments.
The note details specific experimental conditions and analysis parameters for detecting six regulated phthalates (DBP, BBP, DEHP, DNOP, DINP, DIDP) using GC/MS and LC/MS. It provides instructions on sample preparation techniques, such as solid phase extraction for beverages and solvent extraction for food powders, to ensure accurate results that meet regulatory limits.
The maximum allowable limit is 1 part per million (ppm).
A glass cartridge is loaded with 5g of Agilent Chem Elut material.
The run time is 10 minutes.