Agilent DB-5ms Ultra Inert GC Column PAH Analysis Application Note

This application note details a quick and effective analytical method for the determination of low and trace level polycyclic aromatic hydrocarbons (PAHs) in fish samples as an alternative to procedures involving more time consuming and complex techniques. A simplified QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) method with dispersive solid phase extraction (dSPE) provided sufficient sample matrix cleanup while preserving low level analyte detection. A pressure controlled tee was installed post column to implement the use of automated backflush to diminish residual sample carryover and reduce instrument cycle times. The Agilent J&W DB-5ms Ultra Inert (UI) 20 m × 0.18 mm , 0.18 µm column effectively resolved the 16 targeted PAHs. A GC/MS method employing selective ion monitoring (SIM) was calibrated at the 10, 25, 50, 100, 250, 500, and 1000 ng/mL PAH levels, yielding excellent linearity and reproducibility. The spiked levels for recovery determinations were 25, 250, and 500 ng/mL. Recoveries ranged from 80% and 139% with RSD below 6%.


Agilent DB-5ms Ultra Inert GC Column PAH Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
695 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
DB-5MS
MD5 checksum
50B6550B55051BF905675EC602B6FD46

About this manual

This document details an analytical method for determining low and trace level polycyclic aromatic hydrocarbons (PAHs) in fish samples. The method utilizes Agilent Bond Elut QuEChERS dispersive solid phase extraction (dSPE) for sample preparation, offering a simplified alternative to more complex and time-consuming techniques. The approach allows for the processing of up to 60 samples in an eight-hour shift.

The analysis employs an Agilent J&W DB-5ms Ultra Inert (Ul) 20 m x 0.18 mm, 0.18 µm column to effectively resolve the 16 targeted PAHs. The method uses a GC/MS system equipped with selective ion monitoring (SIM) and automated backflush capabilities to reduce residual sample carryover and instrument cycle times.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Chromatographic Conditions
  • Flow Path Supplies
  • Reagents and Chemicals
  • Sample Preparation
  • Discussion of Results
  • Calibration Standards

Specifications

Carrier GasHelium, constant flow 1.7 mL/min
Injection Mode0.5 µL splitless
Oven Program50 °C (0.4 min), 25 °C/min to 195 °C (1.5 min), 8 °C/min to 265 °C, 20 °C/min to 315 °C (1.25 min)
Method Limit of Quantitation10 ppb
Calibration Range10 to 1000 ng/mL

Frequently asked questions

What sample preparation method is used for PAH analysis in fish?

The method uses a QuEChERS approach followed by dispersive solid phase extraction (dSPE).

How many samples can be processed in one shift?

Up to 60 samples can be processed in one eight-hour shift.

What is the limit of quantitation for benzo[a]pyrene?

The method limit of quantitation is 10 ppb for benzo[a]pyrene.