Analysis of Polycyclic Aromatic Hydrocarbons in Fish with Agilent Bond Elut QuEChERS AOAC Kit and HPLC-FLD Application N

An HPLC-Florescence Detection (FLD) method was developed and validated for the determination of sixteen polycyclic aromatic hydrocarbons (PAHs) in fish fillets. The method employs a quick, easy, cheap, effective, rugged and safe (QuEChERS) multiresidue sample preparation procedure adopted from the Association of Analytical Communities (AOAC) Official method 2007.01 for extraction and cleanup. The analytes were separated on an Agilent ZORBAX Eclipse PAH HPLC column (4.6 mm × 50 mm, 1.8 µm) by gradient elution with a binary system of acetonitrile - water and subse- quent fluorescence detection set at appropriate excitation and emission wavelengths. The analyte recoveries ranged from 83.4% to 101% with relative standard deviations ranging from 0.6 to 1.9% at three different fortification levels. The limits of detection and quantification ranged from 0.04 to 0.84 and 0.1 to 2.80 ng/g, respectively.


Analysis of Polycyclic Aromatic Hydrocarbons in Fish with Agilent Bond Elut QuEChERS AOAC Kit and HPLC-FLD Application N - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
433 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
C944547D1E7F675615BCABB0D66C0767

About this manual

The document presents an analytical method developed and validated by researchers at Rhodes University for detecting polycyclic aromatic hydrocarbons (PAHs) in fish tissue. The study utilizes High-Performance Liquid Chromatography with Fluorescence Detection (HPLC-FLD). Sample preparation involves the Agilent Bond Elut QuEChERS AOAC Extraction kit and Dispersive SPE kit to extract and clean up the analytes from fish fillets.

The method was validated by analyzing spiked samples at three different fortification levels. Results showed that analyte recoveries ranged from 83.4% to 101%, with relative standard deviations between 0.6% and 1.9%. The limits of detection (LOD) and quantification (LOQ) for the sixteen target PAHs were determined, ranging from 0.04 ng/g to 0.84 ng/g and 0.1 ng/g to 2.80 ng/g, respectively.

What's inside

  • Abstract
  • Introduction
  • Reagents and Chemicals
  • Standard Solutions
  • Equipment and Material
  • Instrument conditions
  • Extraction and cleanup
  • Results and Discussion
  • Linearity, Limit of Detection (LOD) and Limit of Quantification (LOQ)

Specifications

Flow rate0.8 mL/min
Column temperature18 °C
Injection volume1000 pL (1 mL)
Sample mass5.0 g
Centrifugation speed4000 rpm
Recovery range83.4% to 101%

Frequently asked questions

What compounds were analyzed in the fish samples?

Sixteen polycyclic aromatic hydrocarbons (PAHs) including naphthalene, acenaphthylene, acenaphthene, fluorene, phenanthrene, anthracene, fluoranthene, pyrene, 1,2-benza[a]anthracene, chrysene, benzo[e]pyrene, benzo[e]acenaphthylene, benzo[k]fluoranthene, dibenzo[a,h]anthracene, benzo[g,h,i]perylene and indeno[1,2,3-cd]pyrene.

What is the limit of detection for Fluoranthene?

The limit of detection for Fluoranthene is 0.04 ng/g.

What is the limit of quantification for Dibenzo[a,h]anthracene?

The limit of quantification for Dibenzo[a,h]anthracene is 2.80 ng/g.