Agilent J&W DB-35ms Ultra Inert GC Column Application Note for Organophosphorus Pesticides in Apple Matrix

A method for the determination of low ppm and trace level organophosphorus pesticides residues in apple extract using Agilent J&W DB-35ms Ultra Inert GC Column for GC/MS/FPD


Agilent J&W DB-35ms Ultra Inert GC Column Application Note for Organophosphorus Pesticides in Apple Matrix - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
841 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
DB-35MS
MD5 checksum
62B94BB2093B1DC120721177F3D51209

About this manual

The document describes an analytical method for determining low-level organophosphorus (OP) pesticide residues in apple matrix. It utilizes a capillary flow technology (CFT) device installed post-column to split the effluent between a Mass Selective Detector (MSD) and a Flame Photometric Detector (FPD). This configuration allows for full scan, selective ion monitoring (SIM), and phosphorus-specific detection from a single injection.

The method employs a simplified QuEChERS extraction procedure for sample preparation. The analysis is performed using an Agilent 7890 GC system equipped with an Agilent J&W DB-35ms Ultra Inert column, achieving recovery rates greater than 75% and detection limits between 15-25 ng/mL.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Chromatographic Conditions
  • Flow Path Supplies
  • Reagents and Chemicals
  • Sample Preparation

Frequently asked questions

What is the minimum sample weight required for the apple extraction procedure?

The procedure requires weighing 15 grams of apple sample, with a tolerance of plus or minus 0.1 grams.

What are the specific centrifugation speeds and durations used in the sample preparation steps?

The samples are centrifuged at 4000 rpm for 5 minutes initially, and then at 4000 rpm for 3 minutes after the dispersive SPE step.

What is the split ratio used to divide the column effluent between the MSD and FPD detectors?

The effluent is split in a 3:1 ratio, directing three parts to the MSD and one part to the FPD.