Agilent 720/730 Series ICP-OES Ultra-fast Trace Element Analysis Application Note

This application note describes the implementation of US EPA 200.7 guidelines for the analysis of trace elements in water using the Agilent SVS 2, a new and innovative sample introduction system for the Agilent 720/730 Series ICP-OES.


Agilent 720/730 Series ICP-OES Ultra-fast Trace Element Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
255 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
B534AA95B4CCE9D4A0164440BA8C479E

About this manual

This document details a method for determining trace elements in water conforming to US EPA 200.7 guidelines. It utilizes the Agilent SVS 2, a next-generation sample introduction system designed for the Agilent 720/730 Series ICP-OES to improve sample throughput and reduce operating costs. The SVS 2 maintains a constant flow of solution to the plasma, eliminating the need for long stabilization delays required by conventional systems.

The analysis was performed using an Agilent 720 Series simultaneous ICP-OES with an axially viewed plasma and SPS 3 Sample Preparation System. The document provides specific operating parameters for the instrument and the SVS 2, lists the calibration standards and quality control solutions prepared according to EPA 200.7, and presents the resulting linear dynamic ranges and method detection limits for 32 elements.

What's inside

  • Introduction
  • Instrumentation
  • ICP OES instrument operating parameters
  • Agilent SVS 2 operating parameters
  • Preparation of calibration solutions
  • US EPA 200.7 guidelines summary
  • Linear range analysis
  • Method detection limits
  • Initial OCS performance

Specifications

Plasma gas flow15 L/min
Auxiliary gas flow1.5 L/min
Nebulizer flow0.7 L/min
Pump speed12 rpm
Sample uptake delay time12s
Stabilization time0s
Replicate read time20s
Number of replicates2

Frequently asked questions

What is the stabilization time required when using the SVS 2?

A stabilization of less than 10 seconds is sufficient to load the sample loop and inject the sample into the plasma to attain a stable signal.

How does the SVS 2 improve sample throughput?

The SVS 2 improves sample introduction efficiency by greatly reducing sample uptake and washout times, which can more than double sample throughput.

Which elements are analyzed in this method?

The method analyzes 32 elements in drinking and waste waters, including Ag, Al, As, B, Ba, Be, Ca, Cd, Ce, Co, Cr, Cu, Fe, K, Li, Mg, Mn, Mo, Na, Ni, P, Pb, Se, Si, Sn, Sr, Ti, Tl, V, and Zn.