Agilent LC/MS of Malachite Green Crystal Violet in Fish with Agilent Bond Elut PCX Poroshell 120

A method for simultaneous determination of malachite green and crystal violet, and their metabolites leuco-malachite green and leuco-crystal violet in fish, was developed and validated. The analytes were extracted by solid phase extraction and quantified by liquid chromatography coupled to electrospray ionization tandem mass spectrometry operating in positive ion multiple reaction monitoring mode. The method provided a 0.5 ng/g limit of quantification for all these compounds in fish. The dynamic calibration ranges for these compounds were obtained from 0.5 to 100 ng/g. The overall recoveries ranged from 96 to 109% with RSD values between 1.7 and 4.5%.


Agilent LC/MS of Malachite Green Crystal Violet in Fish with Agilent Bond Elut PCX Poroshell 120 - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
831 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
112E1791D395490038D9B02F8C56A776

About this manual

The document presents a developed and validated analytical method for the simultaneous determination of malachite green (MG), crystal violet (CV), and their metabolites leuco-malachite green (LMG) and leuco-crystal violet (LCV) in fish tissue. The analytes were extracted using solid phase extraction with Agilent Bond Elut Plexa PCX cartridges and quantified via liquid chromatography coupled to electrospray ionization tandem mass spectrometry (LC/MS/MS).

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Equipment and materials
  • Sample preparation
  • Instrument conditions
  • Results and Discussion

Frequently asked questions

What is the limit of quantification for malachite green and crystal violet in fish?

The method provides a 0.5 ng/g limit of quantification for all tested compounds.

What is the dynamic calibration range for the analytes?

The dynamic calibration ranges were obtained from 0.5 to 100 ng/g.

What are the overall recovery percentages and RSD values?

Overall recoveries ranged from 96 to 109% with RSD values between 1.7 and 4.5%.