Agilent 1260 Infinity Bio-inert Quaternary LC System Application Note

This application note demonstrates sensitive and reproducible HPLC analysis of N-linked glycans of human immunoglobulin G using the Agilent 1260 Infinity Bio-inert Quaternary LC System with fluorescence detection. The precision of retention times and areas was determined for analysis of a N-linked human IgG glycan library. In addition, good linearity in a range from 0.008 to 1 pmol was determined for a mix of five glycan standards (Man5, G2, G2S1, G2FS1, and G2S2) with low limits of detection and quantification. The Agilent 1260 Infinity Bio-inert Quaternary LC System with fluorescence detection is an optimal system for sensitive and reproducible analysis of 2-AB derivatized N-glycans released from human immunoglobulin G.


Agilent 1260 Infinity Bio-inert Quaternary LC System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.4 MB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
3A7E6EDA41AD846B1C17B335E9F0B34B

About this manual

This Application Note demonstrates the use of the Agilent 1260 Infinity Bio-inert Quaternary LC System combined with fluorescence detection for the sensitive and reproducible analysis of N-linked glycans derived from human immunoglobulin G (IgG). The study focuses on the separation of 2-AB derivatized N-glycans released from human IgG using Hydrophilic Interaction Chromatography (HILIC)-HPLC.

The document details the experimental setup, including specific Agilent modules such as the Bio-inert Quaternary Pump, High Performance Bio-inert Autosampler, Thermostat, and Fluorescence Detector. It presents results on the precision of retention times and peak areas, the linearity of the method over a range of 0.008 to 1 pmol, and the limits of detection and quantification for a mix of five glycan standards.

What's inside

  • Introduction
  • Experimental
  • Results and discussion
  • Precision of retention time and area
  • Linearity
  • Limit of detections
  • Conclusions
  • References

Specifications

Excitation wavelength260 nm
Emission wavelength430 nm
Column temperature50 °C
Retention time precisionbelow 0.16%
Area precision range0.66% to 4.11%
Limit of detection (LOD)10 to 15 fmol
Limit of quantification (LOQ)34 to 51 fmol

Frequently asked questions

What is the optimal excitation and emission wavelength for this analysis?

The optimized method uses an excitation wavelength of 260 nm and an emission wavelength of 430 nm.

What is the range for linearity determination?

Linearity was determined in a range from 0.008 to 1 pmol using a mix of five glycan standards.