Agilent 6490 Triple Quadrupole LC/MS Application Note for Steroid Detection in Wastewaters

This document introduces the method of optimizing the method parameters of detecting steroids in wastewater using Agilent 6490 Triple Quadrupole LC/MS System with iFunnel Technology, resulting in a sensitivity as low as sub-1 ng/L (1 part per trillion).


Agilent 6490 Triple Quadrupole LC/MS Application Note for Steroid Detection in Wastewaters - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
2.9 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
C74FA8B2EBA329DA5D5C43FA2B0C9AFC

About this manual

This document describes a method for optimizing the detection and quantification of steroids in wastewater using the Agilent 6490 Triple Quadrupole LC/MS System. The study addresses the need for sensitive analysis to monitor environmental contaminants like pharmaceuticals that are not fully removed by sewage treatment plants.

The note provides specific experimental procedures, including sample preparation, cleanup via solid-phase extraction and normal phase chromatography, and detailed instrument conditions. Key optimization parameters discussed include Collision Cell Accelerator Voltage (CAV) and Multiplier Voltage (Delta EMV), which were adjusted to maximize ion response and achieve detection limits as low as sub-1 ng/L.

What's inside

  • Abstract
  • Introduction
  • Reagents and Standards
  • Instruments
  • Sample Collection, Preparation and Cleanup
  • LC and MS Instrument Conditions
  • Results and Discussion

Frequently asked questions

What detection limits were achieved for steroids in wastewaters?

Detection limits as low as sub-1 ng/L (one part per trillion) were demonstrated.

What is the optimized Collision Cell Accelerator Voltage (CAV)?

The optimized CAV was found to be 2 volts, which significantly increased response compared to 5 volts.

What is the optimal range for the Multiplier Voltage (Delta EMV)?

The optimized value was found to be around 200-300 volts.