Agilent 6550 iFunnel Q-TOF LC/MS System Application Note

The document describes the use of the Agilent 6550 iFunnel Q-TOF LC/MS system for high-throughput, high-efficiency metabolome profiling. Compared to the previous generation Q-TOF instrument, this platform has increased sensitivity and resolution, resulting in the detection of an order of magnitude more ions. It provides a 6 to 7-fold increase in coverage of the E. coli metabolome and more detailed pathway analysis.


Agilent 6550 iFunnel Q-TOF LC/MS System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
2.1 MB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
016583643E9E8F89C8B8CA6CCE14F18C

About this manual

This document describes a method for high-throughput, high-efficiency metabolome profiling using the Agilent 6550 iFunnel Q-TOF LC/MS System coupled with flow injection analysis (FIA). The study analyzes small molecules in biological extracts, specifically E. coli, to demonstrate the system's superior performance compared to previous generation instruments.

The application note details the experimental setup, including the use of an Agilent 1290 Series LC pump and Wellplate Autosampler. It presents results showing that the 6550 system detects an order of magnitude more ions than the 6520 platform, leading to a six to seven-fold increase in coverage of the E. coli metabolome and enabling MS/MS identification at concentrations as low as 0.006 pM.

What's inside

  • Introduction
  • Standards and reagents
  • Instruments
  • Sample preparation
  • Analysis of pure standards
  • Analysis of E. coli extracts
  • Metabolite coverage

Specifications

Resolving power40,000
Mass accuracy< 1 ppm
Detection limit0.006 pM
Flow rate150 uL/min
Drying gas temperature325 C
Capillary voltage3,500 V
Fragmentor voltage360 V
Skimmer voltage65 V
Oct RF Vpp750 V
Acquisition range50-1,000 m/z

Frequently asked questions

What is the detection limit for MS/MS identification?

The system enables MS/MS identification of compounds at concentrations as low as 0.006 pM.

How does the 6550 system compare to the 6520 in terms of ion detection?

The 6550 system detects an order of magnitude more ions than the 6520 platform.

What is the increase in metabolome coverage with the 6550 system?

The 6550 system provides a six to seven fold increase in coverage of the E. coli metabolome.